Haemodynamic performance and clinical outcomes of transcatheter aortic valve replacement with the self-expanding ACURATE neo2 (2024)

  • Journal List
  • EuroIntervention
  • v.18(10); 2022 Nov
  • PMC9725034

As a library, NLM provides access to scientific literature. Inclusion in an NLM database does not imply endorsem*nt of, or agreement with, the contents by NLM or the National Institutes of Health.
Learn more: PMC Disclaimer | PMC Copyright Notice

Haemodynamic performance and clinical outcomes of transcatheter aortic valve replacement with the self-expanding ACURATE neo2 (1)

Link to Publisher's site

EuroIntervention. 2022 Nov; 18(10): 804–811.

Published online 2022 Nov 18. doi:10.4244/EIJ-D-22-00289

PMCID: PMC9725034

PMID: 35678222

Andrea Scotti, MD, Matteo Pagnesi, MD, Won-Keun Kim, MD, Ulrich Schäfer, MD, Marco Barbanti, MD, Giuliano Costa, MD, Sara Baggio, MD, Matteo Casenghi, MD, Federico De Marco, MD, Maarten Vanhaverbeke, MD,PhD, Lars Sondergaard, MD, Alexander Wolf, MD, Joachim Schofer, MD, Marco Bruno Ancona, MD, Matteo Montorfano, MD,MD, Ran Kornowski, MD,MD, Hana Vaknin Assa, MD,MD, Stefan Toggweiler, MD,MD, Alfonso Ielasi, MD,MD, David Hildick-Smith, MD,MD, Stephan Windecker, MD,MD, Albrecht Schmidt, MD,MD, Andrea Buono, MD,MD, Diego Maffeo, MD,MD, Dimytri Siqueira, MD,MD, Francesco Giannini, MD,MD, Marianna Adamo, MD,MD, Mauro Massussi, MD,MD, David A. Wood, MD,MD, Jan-Malte Sinning, MD,MD, Jan Van Der Heyden, MD,MD, Dirk-Jan van Ginkel, MD,MD, Nicholas Van Mieghem, MD,MD, Verena Veulemans, MD,MD, Darren Mylotte, MD,MD, Vasileios Tzalamouras, MD,MD, Maurizio Taramasso, MD,MD, Rodrigo Estévez-Loureiro, MD,MD,PhD,PhD, Antonio Colombo, MD,MD, Antonio Mangieri, MD, and Azeem Latib, MD,MD*

Author information Article notes Copyright and License information PMC Disclaimer

Associated Data

Supplementary Materials

Abstract

Background

Transcatheter aortic valve replacement (TAVR) with the ACURATE neo device has been associated with a non-negligible incidence of paravalvular aortic regurgitation (AR). The new-generation ACURATE neo2 has been designed to mitigate this limitation.

Aims

The aim of the study was to compare TAVR with the ACURATE neo and neo2 devices.

Methods

The NEOPRO and NEOPRO-2 registries retrospectively included patients undergoing transfemoral TAVR with self-expanding valves at 24 and 20 centres, respectively. Patients receiving the ACURATE neo and neo2 devices (from January 2012 to December 2021) were included in this study. Predischarge and 30-day VARC-3 defined outcomes were evaluated. The primary endpoint was predischarge moderate or severe paravalvular AR. Subgroup analyses per degree of aortic valve calcification were performed.

Results

A total of 2,026 patients (neo: 1,263, neo2: 763) were included. Predischarge moderate or severe paravalvular AR was less frequent for the neo2 group (2% vs 5%; p<0.001), resulting in higher VARC-3 intended valve performance (96% vs 90%; p<0.001). Furthermore, more patients receiving the neo2 had none/trace paravalvular AR (59% vs 38%; p<0.001). The reduction in paravalvular AR with neo2 was mainly observed with heavy aortic valve calcification. New pacemaker implantation and VARC-3 technical and device success rates were similar between the 2 groups; there were more frequent vascular and bleeding complications for the neo device. Similar 1-year survival was detected after TAVR (neo2: 90% vs neo: 87%; p=0.14).

Conclusions

TAVR with the ACURATE neo2 device was associated with a lower prevalence of moderate or severe paravalvular AR and more patients with none/trace paravalvular AR. This difference was particularly evident with heavy aortic valve calcification.

Introduction

Transcatheter aortic valve replacement (TAVR) is an established treatment option for patients with symptomatic severe aortic stenosis1,2. As TAVR candidates are increasingly younger and at lower surgical risk, it has become crucial to minimise potential procedural complications and to provide surgical-like long-term outcomes. Post-procedural moderate or severe paravalvular aortic regurgitation (AR) is a relevant complication after TAVR that has been found to be associated with adverse short- and long-term outcomes3. In the last few years, the first-generation ACURATE neo (Boston Scientific) transcatheter heart valve (THV) has emerged as a widely adopted self-expanding device for TAVR, associated with good procedural and clinical outcomes4,5,6,7,8,9. However, 2 randomised trials have recently reported a higher rate of moderate or severe paravalvular AR with the ACURATE neo as compared to other new-generation self-expanding and balloon-expandable THVs10,11. This complication was more frequent with increased device landing zone calcification5. For this reason, careful patient selection, proper sizing, and appropriate positioning were proposed to optimise procedural outcomes9.

In September 2020, the new-generation ACURATE neo2 THV was commercially released in Europe. This latest iteration of the ACURATE neo platform has been specifically designed to minimise the occurrence of paravalvular AR by utilising a 60% larger sealing skirt. Quantitative aortographic assessments have shown promising results in terms of paravalvular AR reduction with the ACURATE neo2 device12. However, no large, real-world data have compared the performance of ACURATE neo and neo2. With this background, our study aimed to investigate the haemodynamic performance and clinical outcomes after transfemoral TAVR with the ACURATE neo2 compared to the first-generation ACURATE neo THV.

Methods

Study population

The observational, retrospective NEOPRO (A Multicenter Comparison of Acurate NEO Versus Evolut PRO Transcatheter Heart Valves) registry included a total of 1,551 patients who underwent transfemoral TAVR with either ACURATE neo (n=1,263) or Evolut PRO (n=288; Medtronic) devices between January 2012 and March 2018 at 24 centres4. The NEOPRO-2 registry was designed to expand the previous registry to include procedures performed with the new-generation ACURATE neo2 (n=763) and Evolut PRO or PRO+ (n=1,412) devices up to December 2021 at 20 centres. All consecutive patients treated with transfemoral TAVR for symptomatic, severe aortic stenosis of the native aortic valve (AV) with the implantation of the aforementioned self-expanding devices were included in the registries. For the purposes of the present study, only patients treated with ACURATE neo or neo2 THVs were analysed. The number of patients included from each participating centre is detailed in Supplementary Table 1. The treatment period was from January 2012 to March 2018 for the neo THV and from September 2020 to December 2021 for the neo2 device. Data obtained from 29 participating centres were included in the present analysis: 18 centres implanting the ACURATE neo (NEOPRO) and 16 centres using the neo2 device (NEOPRO-2). Local multidisciplinary Heart Teams evaluated each case and confirmed eligibility for transfemoral TAVR. All patients provided written informed consent for the procedure and subsequent data collection per local practice for retrospective data. The study complied with the Declaration of Helsinki and was approved by local ethics committees. Preprocedural screening was performed by means of clinical assessment (patient demographics, symptoms, comorbidities, laboratory examinations, and risk evaluation), echocardiography and multidetector computed tomography.

Device description

The ACURATE neo2 bioprosthesis is a self-expanding THV with a supra-annular leaflet design. Three sizes are currently available (small, medium, and large) and correspond to annular diameters of 21-23, 23-25, and 25-27 mm, respectively. The neo2 THV is implanted using a dedicated transfemoral delivery system inserted through a 14 Fr expandable sheath (iSleeve [Boston Scientific]). The deployment is performed in a top-down sequence, starting with the release of the stabilisation arches, and does not require rapid pacing. The self-expanding nitinol frame is wrapped with a pericardial sealing skirt on the outer and inner surface of the stent body that extends 60% higher from the inflow part of the stent frame as compared to the first-generation ACURATE neo. The skirt’s extended dimensions have the potential of providing a more synchronous adaptation to the native aortic annulus during the different phases of the cardiac cycle, especially in irregular and calcified anatomies. Furthermore, a radiopaque marker has been added to the delivery system to navigate accurate positioning of the THV at the aortic annulus.

Study endpoints and definitions

The primary study endpoint was the occurrence of moderate or severe paravalvular AR at predischarge transthoracic echocardiography. Secondary endpoints were Valve Academic Research Consortium (VARC)-3 defined clinical outcomes at 30 days13, including the need for permanent pacemaker implantation (PPI), and 1-year overall survival.

Paravalvular AR severity was assessed with Doppler echocardiography according to VARC-3 criteria and classified as follows: none or trace, mild, moderate, and severe13. Native aortic valve and left ventricular outflow tract calcifications from multidetector computed tomography scans were classified and graded using a semiquantitative scoring system, as previously described6,14.

Statistical analysis

Continuous variables are reported as mean±standard deviation or median (interquartile range [IQR]) and compared with the Student’s unpaired t-test (parametric test) or the Wilcoxon rank-sum test (non-parametric test), according to their distribution. Categorical variables were reported as absolute and relative frequencies and compared with the χ2 test with Yates’ correction for continuity or Fisher’s exact test, as appropriate. Survival curves with their 95% confidence interval (CI) were plotted using the Kaplan-Meier estimator and compared with the log-rank test. A subgroup analysis testing the primary and secondary endpoints across different degrees of aortic valve calcification was also performed. For all analyses, a two-sided p-value <0.05 was considered to be significant. Statistical analyses were performed using R, version 4.0.2 (R Foundation).

Results

Baseline patient characteristics

A total of 2,026 patients in the NEOPRO and NEOPRO-2 registries underwent transfemoral TAVR with the self-expanding ACURATE neo or neo2 THVs and were included in this study. Of these, 1,263 patients received the first-generation ACURATE neo, whereas 763 were treated using the new ACURATE neo2 device. Baseline characteristics of the study population are reported in Table 1. The mean age was 82.0±5.8 years, and 66% of patients were women. Patients treated with the ACURATE neo were more frequently in New York Heart Association (NYHA) Functional Class III or IV (78% vs 55%; p<0.001), and were deemed at higher surgical risk (STS score: 4.1 [IQR 2.9-6.1] vs 3.5 [IQR 2.5-5.0], EuroSCORE II: 4.4 [IQR 2.7-7.2] vs 3.1 [IQR 2.1-5.1]; p<0.001). Patients receiving the ACURATE neo2 were characterised by smaller anatomies of the aortic valve (area and perimeter) and of the TAVR femoral access (minimal diameter: 7.14±1.13 vs 7.95±1.37; p<0.001). The severity of aortic valve calcification was higher in the neo group, whereas mild and moderate left ventricular outflow tract calcifications were more frequent in neo2 patients.

Table 1

Baseline characteristics.

Total
(2,026)
ACURATE neo (1,263)ACURATE neo2 (763)p-value
Clinical characteristics
Age, years82±5.882±5.882±5.90.822
Male694 (34)444 (35)250 (33)0.303
BMI27±527±527±50.281
BSA1.82±0.211.82±0.211.82±0.220.314
Hypertension1,726 (87)1,079 (88)647 (85)0.055
Diabetes mellitus598 (30)379 (30)219 (29)0.571
Atrial fibrillation648 (32)408 (33)240 (32)0.772
Previous stroke209 (10)126 (10)83 (11)0.719
Peripheral vascular disease271 (13)156 (12)115 (15)0.093
Previous myocardial infarction220 (11)138 (12)82 (11)0.689
Previous PCI582 (29)370 (29)212 (28)0.496
Previous CABG194 (10)147 (12)47 (6)<0.001
COPD369 (18)244 (19)125 (16)0.114
eGFR, ml/min/1.73m260±2558±2264±29<0.001
Prior PM/ICD219 (11)158 (12)61 (8)0.002
NYHA Class III/IV1,397 (69)981 (78)416 (55)<0.001
EuroSCORE II3.9 [2.5-6.6]4.4 [2.7-7.2]3.1 [2.1-5.1]<0.001
STS score (mortality)4.0 [2.8-5.8]4.1 [2.9-6.1]3.5 [2.5-5.0]<0.001
Echocardiographic data
Mean aortic gradient, mmHg44±1643±1745±140.057
AVA, cm20.71±0.220.71±0.190.71±0.260.523
Indexed AVA, cm2/m20.39±0.100.39±0.110.39±0.090.891
LVEF, %57±1157±1258±100.032
CT analysis
Aortic valve area, mm2429±64432±67424±590.015
Aortic valve perimeter, mm74±675±674±5<0.001
Aortic valve calcificationNone or mild470 (29)286 (29)184 (29)0.02
Moderate739 (46)428 (44)311 (50)
Heavy396 (25)263 (27)133 (21)
LVOT calcificationNone693 (53)541 (56)152 (46)0.01
Mild379 (29)265 (27)114 (35)
Moderate152 (12)107 (11)45 (14)
Severe78 (6)62 (6)16 (5)
Femoral access*, mm7.78±1.367.95±1.377.14±1.13<0.001
*Minimal lumen diameter of the femoral artery used to deliver the valve. Values are n (%), mean±standard deviation, or median [interquartile range]. AVA: aortic valve area; BMI: body mass index; BSA: body surface area; CABG: coronary artery bypass grafting; COPD: chronic obstructive pulmonary disease; CT: computed tomography; eGFR: estimated glomerular filtration rate; ICD: implantable cardioverter defibrillator; LVEF: left ventricular ejection fraction; LVOT: left ventricular outflow tract; NYHA: New York Heart Association; PCI: percutaneous coronary intervention; PM: pacemaker; STS: Society of Thoracic Surgeons

Open in a separate window

Procedural results

Procedural characteristics are depicted in Table 2. Valve sizes were equally distributed within the 2 groups. Implantation of the neo THV was more frequently performed under general anaesthesia (13% vs 1%; p<0.001) and completed with final post-dilatation (42% vs 31%; p<0.001). The prevalence of procedural complications (namely: death, valve embolisation, the need for a second THV, annular rupture, pericardial tamponade, aortic dissection, coronary occlusion, and conversion to open-heart surgery) was low with no differences between the neo and neo2 groups.

Table 2

Procedural characteristics.

Total
(2,026)
ACURATE neo (1,263)ACURATE neo2 (763)p-value
Valve size, mm23533 (26)348 (28)185 (24)0.255
25847 (42)520 (41)327 (43)
27645 (32)394 (31)251 (33)
General anaesthesia175 (9)168 (13)7 (1)<0.001
Predilatation1,706 (84)1,051 (83)655 (86)0.141
Post-dilatation759 (38)526 (42)233 (31)<0.001
Death2 (0.3)0 (0.0)2 (0.3)1.000
Valve embolisation21 (1)13 (1)8 (1)1.000
Second THV implanted20 (1)14 (1)6 (1)0.634
Annular rupture5 (0.2)4 (0.3)1 (0.1)0.656
Pericardial tamponade27 (1)20 (2)7 (1)0.286
Aortic dissection1 (0.0)1 (0.1)0 (0.0)1.000
Coronary occlusion4 (0.2)2 (0.2)2 (0.3)1.000
Conversion to open-heart surgery16 (1.0)13 (1.0)3 (0.4)0.191
Values are expressed as n (%). THV: transcatheter heart valve

Open in a separate window

Clinical outcomes

Clinical outcomes assessed at 30 days post-TAVR are shown in Table 3 and the Central illustration. VARC-3 defined technical (neo: 91% vs neo2: 93%; p=0.117) and device success (neo: 81% vs neo2: 84%; p=0.119) were similar in the 2 groups. The VARC-3 defined intended performance of the valve was more frequently met in the neo2 group (96% vs 90%; p<0.001). This result was mainly driven by a lower rate of moderate or severe paravalvular AR in the neo2 group (2% vs 5%; p<0.001). The prevalence of none/trace paravalvular AR was significantly higher after TAVR with the neo2 THV (59% vs 38%; p<0.001). Slightly increased mean aortic gradients were found in the neo2 group (neo: 8.0±3.3 mmHg vs neo2: 8.9±4.1 mmHg; p<0.001), and no differences were found in terms of aortic valve area. The need for a new PPI was similar in the 2 groups (neo: 9% vs neo2: 8%; p=0.46). Patients receiving the neo2 THV experienced fewer vascular (p<0.001) and bleeding (p=0.020) complications.

Open in a separate window

Central Illustration

Comparison of the ACURATE neo and the ACURATE neo2 THVs.

A) Device characteristics, B) VARC-3 technical success, C) predischarge paravalvular aortic regurgitation, and D) the need for permanent pacemaker implantation after transcatheter aortic valve replacement. Device illustrations reproduced with permission from Boston Scientific. CSC: cardiac structural complication; THV: transcatheter heart valve; VARC: Valve Academic Research Consortium

Table 3

30-day outcomes.

Total
(2,026)
ACURATE neo (1,263)ACURATE neo2 (763)p-value
All-cause death61 (3)39 (3)22 (3)0.903
VARC 3 – technical success1,859 (92)1,149 (91)710 (93)0.117
VARC 3 – device success1,630 (82)1,024 (81)606 (84)0.119
VARC 3 – intended performance1,286 (93)572 (90)714 (96)<0.001
PM implantation147 (8)96 (9)51 (8)0.460
Acute kidney injury (stage 2-3)58 (3)37 (3)21 (3)0.953
Vascular complicationsNone1,700 (87)1,032 (83)668 (94)<0.001
Minor156 (8)138 (11)18 (2)
Major98 (5)75 (6)23 (3)
Bleeding complicationsNone1,638 (86)1,011 (85)627 (88)0.020
Type 1104 (6)65 (6)39 (6)
Type 279 (4)56 (5)23 (3)
Type 374 (4)56 (5)18 (2)
Type 42 (0.1)0 (0.0)2 (0.3)
Mean aortic gradient, mmHg8.5±3.88±3.38.9±4.1<0.001
AVA, cm21.8±0.41.8±0.41.8±0.40.826
Indexed AVA, cm2/m20.96±0.200.96±0.200.96±0.200.769
Moderate or severe paravalvular AR*75 (4)62 (5)13 (2)<0.001
*Predischarge assessment. Values are n (%) or mean±standard deviation. AR: aortic regurgitation; AVA: aortic valve area; PM: pacemaker; VARC: Valve Academy Research Consortium

Open in a separate window

Patients were followed up for a median time of 83 [IQR 30-261] days. As shown in the Kaplan-Meier analysis (Figure 1), all-cause mortality at 1-year follow-up was not significantly different between the neo2 and neo groups (90%, 95% CI: 83-98 vs 87%, 95% CI: 84-90; p=0.14).

Open in a separate window

Figure 1

Kaplan-Meier analysis of overall survival in ACURATE neo vs ACURATE neo2 transcatheter aortic valves.

Subgroup analysis on aortic valve calcification severity

As shown in Table 4, clinical outcomes were further analysed stratifying the overall population for baseline aortic valve calcification grades (none or mild, moderate, and heavy). The significant reduction of moderate or severe paravalvular AR in the neo2 group was observed in the subgroup of patients with heavy aortic valve calcification (2% vs 9%; p=0.018); consequently, VARC-3 intended performance of the valve was more frequently met among patients receiving neo2 THV in the heavy aortic valve calcification subgroup (97% vs 88%; p=0.005). No significant differences were observed for these 2 endpoints between the neo and neo2 groups in the other calcification subgroups (Table 4).

Table 4

30-day outcomes stratified per aortic valve calcification grade.

None or mild calcificationModerate calcificationHeavy calcificationp-value
for interaction
ACURATE neo
(286)
ACURATE neo2 (184)p-valueACURATE neo
(428)
ACURATE neo2 (311)p-valueACURATE neo
(263)
ACURATE neo2 (133)p-value
All-cause death9 (3)6 (3)1.00011 (3)9 (3)0.9629 (4)0 (0)0.0700.418
VARC 3 - technical success250 (87)167 (91)0.332390 (91)287 (92)0.669240 (91)128 (96)0.1050.404
VARC 3 - device success229 (80)150 (84)0.311346 (81)244 (83)0.524203 (78)100 (85)0.1720.691
VARC 3 - intended performance262 (97)178 (97)0.937377 (93)293 (95)0.298214 (88)127 (97)0.0050.899
PM implantation21 (9)11 (7)0.60540 (10)23 (9)0.53517 (8)6 (6)0.7050.982
Mean aortic gradient, mmHg7.6±3.48.5±4.20.0238.0±3.29.0±4.10.0018.2±3.68.9±3.90.0230.141
AVA, cm21.7±0.41.7±0.40.8671.7±0.41.7±0.40.7311.8±0.41.9±0.40.8670.135
Indexed AVA, cm2/m20.96±0.200.96±0.200.9790.93±0.200.94±0.200.5890.96±0.201.0±0.20.9790.208
Moderate or severe paravalvular AR*6 (2)1 (0.5)0.31721 (5)7 (2)0.07723 (9)3 (2)0.0180.671
*Predischarge assessment. Values are n (%) or mean±standard deviation. AR: aortic regurgitation; AVA: aortic valve area; PM: pacemaker; VARC: Valve Academic Research Consortium

Open in a separate window

All the remaining clinical outcomes were consistent with the primary analysis with no significant differences between both groups across all aortic valve calcification subgroups. In detail, the rates of new PPI were similar between neo and neo2 THVs in each subgroup (none/mild: 9% vs 7%, moderate: 10% vs 9%, heavy: 8% vs 6%; all p-values>0.05, p for interaction=0.982).

Discussion

The main findings of our multicentre, observational, real-world comparison between ACURATE neo and neo2 devices in a total of 2,026 patients undergoing transfemoral TAVR (from the NEOPRO/NEOPRO-2 registries) are as follows: 1) the latest-generation ACURATE neo2 THV was associated with a significant reduction in post-procedural moderate or severe paravalvular AR as compared to the first-generation neo THV; 2) a similar need for new PPI was observed between neo2 and neo devices; 3) the superior performance of the ACURATE neo2 THV was particularly evident in severely calcified aortic valve anatomies; 4) TAVR with the ACURATE neo2 in combination with the expandable iSleeve is associated with reduced rates of vascular complications.

In our study, procedural outcomes after ACURATE neo2 THV implantation suggest acceptable safety and efficacy with rates of 3% for 30-day mortality, 93% for VARC-3 technical success, 84% for VARC-3 device success, and 96% for VARC-3 intended performance of the valve. These results compare favourably with available evidence on the first-generation neo device reporting an equal 30-day mortality rate (3%) and similar rates of procedural complications5,10,11. Interestingly, we observed fewer vascular and bleeding complications in the neo2 group as compared to the neo group. It must be acknowledged that all the procedures were transfemoral with smaller accesses for neo2 recipients (minimal lumen diameter 7.14±1.13 vs 7.95±1.37; p<0.001), similar baseline patient characteristics (age: 82±6 years, atrial fibrillation: 33% vs 32%, peripheral vascular disease: 12% vs 15%), and lower surgical risk scores for the neo2 group, in accordance with TAVR indication expanding to lower-risk patients over time. Even if speculative, the fewer bleeding and vascular complications observed with the implantation of the neo2 THV may be explained by a combination of the redesigned expandable introducer with a low-profile (iSleeve; Boston Scientific), increased operator experience with the ACURATE system, and improved vascular access/complication management.

The only VARC-3 defined outcome that was significantly different between the neo and neo2 devices was the intended performance of the valve (neo2: 96% vs neo: 90%; p<0.001). This composite endpoint is achieved with a mean transvalvular gradient <20 mmHg, peak velocity <3 m/s, Doppler velocity index ≥0.25, and less than moderate AR13. The improved performance of the neo2 THV was driven by a reduction in moderate or severe paravalvular AR compared to the first-generation neo THV. In the overall population, the 5% rate of moderate or severe paravalvular AR that occurred after the ACURATE neo implantation was significantly higher than the 2% rate observed with the neo2 THV (p<0.001). The rate of paravalvular AR in our neo group (5%) is lower than those reported by the randomised SCOPE 1 (9.4%) and SCOPE 2 (9.6%) trials10,11. This finding may be explained by differences in outcome adjudication (core lab vs centre-reported) and baseline patient characteristics. Although these trials excluded severe eccentric aortic valve calcifications, they did not report the degrees of overall calcifications which have been demonstrated by Kim et al as having a significant impact on moderate or severe paravalvular AR rates (mild: 0.8%, moderate: 5%, severe: 13%)5. Accordingly, our subgroup analysis reported a similarly increasing trend for paravalvular AR in the neo group, starting from 2% for none/mild aortic valve calcifications, up to 5% for moderate and 9% for heavy calcifications. On the contrary, the rate of moderate or severe paravalvular AR after TAVR with the neo2 THV was consistent among these 3 calcification subgroups (none/mild: 0.5%, moderate: 2%, heavy: 2%). Similar rates of moderate or severe paravalvular AR (1.7-2.5%) were reported in other exploratory analyses evaluating TAVR with the ACURATE neo2 device12,15. Furthermore, 59% of patients treated with the new-generation neo2 device showed none/trace paravalvular AR after TAVR, which is significantly higher than the 38% obtained with the neo THV (p<0.001). This finding also compares favourably with the frequency of none/trace paravalvular AR observed after neo implantation in the SCOPE 1 (40%) and SCOPE 2 (27%) trials10,11. These results are promising as they demonstrate how the engineering refinements translate into better performance of the ACURATE neo2 THV. Moreover, they indicate that the caveat of avoiding patients with severe aortic calcifications may no longer be appropriate for the ACURATE neo2 system9. Pending further supporting evidence, it seems that this new-generation device can provide favourable performance with a low rate of significant paravalvular AR, even in the more challenging calcific anatomies. The observed gradient with neo2 in our population (8.9±4.1 mmHg) is similar to the one reported in patients with small annuli included in the TAVI-SMALL registry and receiving the neo device (9.6±0.3 mmHg)16. An inverse correlation between annular dimensions and post-procedural gradients has been previously demonstrated with better haemodynamic performance of self-expanding supra-annular THVs in this anatomical setting17,18.

Another relevant clinical outcome after TAVR is represented by the need for new PPI. Previous studies investigating the ACURATE neo system reported a 10-11% rate of PPI5,10,11,which was found to be independent of device landing zone calcification5. Our analysis confirms these findings for the neo2 THV, showing a rate of new PPI equal to 8% with no significant differences across the aortic valve calcification subgroups. These results are even more meaningful when compared with the 17-18% of PPI after TAVR with the self-expanding CoreValve Evolut platform (Medtronic)10,19. A stable and precise (less protrusion in the left ventricular outflow tract) valve implantation with top-down deployment and radiopaque positioning markers, a moderate device radial force, and the temporal shift from the left anterior oblique to cusp overlap view for the THV implantation are the technical factors explaining this relatively low PPI rate. Keeping this figure as low as possible has a great clinical value and may be used to guide device selection in patients at high risk of permanent conduction disturbances after TAVR20.

At 1-year follow-up, we did not observe a significant difference in all-cause mortality between the neo2 and neo groups (10% vs 13%; p=0.14). The 1-year mortality rate observed in this real-world experience with ACURATE neo is in line with rates recently reported by the randomised SCOPE 1 (11%) and SCOPE 2 (13%) trials10,11. Given the prognostic impact of moderate or severe paravalvular AR3 and its significant reduction using the ACURATE neo2, further analyses with larger sample sizes and longer follow-up are eagerly awaited to better explore long-term outcomes after TAVR with this new self-expanding platform.

Limitations

The main limitation of this study is related to its retrospective observational design, with no core laboratory analysis of procedural results and echocardiographic readings or independent adjudication of clinical events. The investigated THVs were implanted in consecutive time periods, and unmeasured confounding factors (e.g., operator experience, valve preference) may have affected the presented results. The use of multiple different sheaths with the ACURATE neo may be a major contributor to the differences in peripheral vascular characteristics compared to the neo2, which was implanted through a redesigned expandable introducer with a low-profile (iSleeve; Boston Scientific). Given the recent release of the ACURATE neo2 device (September 2020), available follow-up time is currently limited, and future studies will be needed to assess if lower paravalvular AR is sustained over time and how it impacts on long-term clinical outcomes. Whilst waiting for the results of ongoing registries (Early neo2 Registry of the Acurate neo2 TAVI Prosthesis [ClinicalTrials.gov: NCT04810195]) and randomised controlled trials (ACURATE IDE: Safety and Efficacy Study of Acurate Valve for Transcatheter Aortic Valve Replacement [ClinicalTrials.gov: NCT03735667]), these exploratory analyses can provide immediate assistance in THV selection for TAVR.

Conclusions

The latest-generation ACURATE neo2 THV is associated with a lower rate of moderate or severe paravalvular AR, when compared to the first-generation ACURATE neo, in patients undergoing transfemoral TAVR. As a result, a greater percentage of patients receiving the neo2 THV have none/trace paravalvular AR. The superior performance of the neo2 device is particularly evident among patients with heavy aortic valve calcification.

Impact on daily practice

TAVR with the ACURATE neo THV was associated with a non-negligible rate of moderate or severe paravalvular AR, which is known to have an adverse prognostic impact. TAVR with the ACURATE neo2 device is associated with a lower rate of moderate or severe paravalvular AR when compared with the first-generation ACURATE neo. The superior performance of the neo2 device is particularly evident among patients with heavy aortic valve calcification. Further studies with larger sample sizes and longer follow-up are needed to confirm these preliminary findings.

Guest editor

This paper was guest edited by Franz-Josef Neumann, MD, Department of Cardiology and Angiology II, University Heart Center Freiburg - Bad Krozingen, Bad Krozingen, Germany.

Supplementary data

Supplementary Table 1:

NEOPRO and NEOPRO-2 participating centres.

Click here to view.(102K, pdf)

Conflict of interest statement

A. Latib has served on advisory boards or as a consultant for Medtronic, Boston Scientific, Philips, Edwards Lifesciences, and Abbott. W-K. Kim is a proctor for Boston Scientific and Abbott Vascular; and has received speaker fees from Edwards Lifesciences. U. Schäfer is proctor for Boston Scientific and Medtronic and has received lecture fees and travel support from both companies. M. Barbanti has served as a consultant for Edwards Lifesciences. D. Hildick-Smith is an advisor and a proctor for Boston Scientific, Symetis, and Medtronic. A. Wolf is a proctor for Medtronic and Boston Scientific. N. Van Mieghem has received research grant support from Boston Scientific, Abbott Vascular, Medtronic, Claret, and Essential Medical. S. Toggweiler is a consultant and proctor for Boston Scientific, Medtronic, Abbott, and Biosensors; is a consultant for Medira, Atheart Medical, Veosource, Shockwave, and Teleflex; has received institutional research grants from Boston Scientific and Fumedica AG; and holds equity in Hi-D Imaging. D. Mylotte is a proctor for Medtronic and Microport. V. Veulemans has received lecture fees and travel support from Medtronic and Edwards Lifesciences. S. Windecker has received grants from Abbott Vascular, Biotronik, Boston Scientific, Edwards Lifesciences, and Medtronic. D. Siqueira is proctor for Medtronic, Symetis, and Edwards Lifesciences. M. Adamo reports speaker honoraria from Abbott Vascular and Medtronic. J-M. Sinning is a proctor for Medtronic and Boston Scientific; and has received speaking honoraria and research grants from Medtronic, Edwards Lifesciences, and Boston Scientific. D.A. Wood has received grant support from Boston Scientific; and is a consultant for Medtronic. L. Sondergaard has received consultant fees and institutional research grants from Boston Scientific. The other authors have no conflicts of interest to declare. The Guest Editor reports lecture fees paid to his institution from Amgen, Bayer Healthcare, Biotronik, Boehringer Ingelheim, Boston Scientific, Daiichi Sankyo, Edwards Lifesciences, Ferrer, Pfizer, and Novartis; consultancy fees paid to his institution from Boehringer Ingelheim; and grant support from Bayer Healthcare, Boston Scientific, Biotronik, Edwards Lifesciences, GlaxoSmithKline, Medtronic, and Pfizer.

Abbreviations

ARaortic regurgitation
PPIpermanent pacemaker implantation
TAVRtranscatheter aortic valve replacement
THVtranscatheter heart valve
VARCValve Academic Research Consortium

Contributor Information

Andrea Scotti, Montefiore Einstein Center for Heart and Vascular Care, Montefiore Medical Center, Albert Einstein College of Medicine, Bronx, NY, USA. Cardiovascular Research Foundation, New York, NY, USA.

Matteo Pagnesi, ASST Spedali Civili di Brescia and Department of Medical and Surgical Specialties, Radiological Sciences and Public Health, University of Brescia, Brescia, Italy.

Won-Keun Kim, Department of Cardiology and Cardiac Surgery, Kerckhoff Heart and Lung Center, Bad Nauheim, Germany.

Ulrich Schäfer, Department of Cardiology, University Heart & Vascular Center, Hamburg, Germany.

Marco Barbanti, Department of Cardiology, C.A.S.T. Policlinic G. Rodolico Hospital, University of Catania, Catania, Italy.

Giuliano Costa, Department of Cardiology, C.A.S.T. Policlinic G. Rodolico Hospital, University of Catania, Catania, Italy.

Sara Baggio, Cardio Center, Humanitas Research Hospital, Rozzano-Milan, Italy.

Matteo Casenghi, Department of Cardiology, IRCCS Policlinico San Donato, Milan, Italy.

Federico De Marco, Department of Cardiology, IRCCS Policlinico San Donato, Milan, Italy.

Maarten Vanhaverbeke, The Heart Centre, Rigshospitalet, Copenhagen University Hospital, Copenhagen, Denmark.

Lars Sondergaard, The Heart Centre, Rigshospitalet, Copenhagen University Hospital, Copenhagen, Denmark.

Alexander Wolf, Contilia Heart and Vascular Centre, Elisabeth-Krankenhaus Essen, Essen, Germany. Contilia Heart and Vascular Centre, Elisabeth-Krankenhaus Essen, Essen, Germany.

Joachim Schofer, Department for Percutaneous Treatment of Structural Heart Disease, Albertinen Heart Center, Hamburg, Germany.

Marco Bruno Ancona, Cardio-Thoracic-Vascular Department, San Raffaele Scientific Institute, Milan, Italy. Cardio-Thoracic-Vascular Department, San Raffaele Scientific Institute, Milan, Italy.

Matteo Montorfano, Cardio-Thoracic-Vascular Department, San Raffaele Scientific Institute, Milan, Italy. Cardio-Thoracic-Vascular Department, San Raffaele Scientific Institute, Milan, Italy.

Ran Kornowski, Department of Cardiology, Rabin Medical Center-Beilinson Hospital, Petah Tikva, Israel. Department of Cardiology, Rabin Medical Center-Beilinson Hospital, Petah Tikva, Israel.

Hana Vaknin Assa, Department of Cardiology, Rabin Medical Center-Beilinson Hospital, Petah Tikva, Israel. Department of Cardiology, Rabin Medical Center-Beilinson Hospital, Petah Tikva, Israel.

Stefan Toggweiler, Department of Cardiology, Heart Center Lucerne, Luzerner Kantonsspital, Lucerne, Switzerland. Department of Cardiology, Heart Center Lucerne, Luzerner Kantonsspital, Lucerne, Switzerland.

Alfonso Ielasi, Clinical and Interventional Cardiology Unit, Istituto Clinico Sant'Ambrogio, Milan, Italy. Clinical and Interventional Cardiology Unit, Istituto Clinico Sant'Ambrogio, Milan, Italy.

David Hildick-Smith, Department of Cardiology, Royal Sussex County Hospital, Brighton, UK. Department of Cardiology, Royal Sussex County Hospital, Brighton, UK.

Stephan Windecker, Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland. Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.

Albrecht Schmidt, Division of Cardiology, Medical University of Graz, Graz, Austria. Division of Cardiology, Medical University of Graz, Graz, Austria.

Andrea Buono, Interventional Cardiology Unit, Fondazione Poliambulanza, Brescia, Italy. Interventional Cardiology Unit, Fondazione Poliambulanza, Brescia, Italy.

Diego Maffeo, Interventional Cardiology Unit, Fondazione Poliambulanza, Brescia, Italy. Interventional Cardiology Unit, Fondazione Poliambulanza, Brescia, Italy.

Dimytri Siqueira, Instituto Dante Pazzanese de Cardiologia, São Paulo, Brazil. Instituto Dante Pazzanese de Cardiologia, São Paulo, Brazil.

Francesco Giannini, GVM Care and Research, Maria Cecilia Hospital, Cotignola, Italy. GVM Care and Research, Maria Cecilia Hospital, Cotignola, Italy.

Marianna Adamo, ASST Spedali Civili di Brescia and Department of Medical and Surgical Specialties, Radiological Sciences and Public Health, University of Brescia, Brescia, Italy. ASST Spedali Civili di Brescia and Department of Medical and Surgical Specialties, Radiological Sciences and Public Health, University of Brescia, Brescia, Italy.

Mauro Massussi, ASST Spedali Civili di Brescia and Department of Medical and Surgical Specialties, Radiological Sciences and Public Health, University of Brescia, Brescia, Italy. ASST Spedali Civili di Brescia and Department of Medical and Surgical Specialties, Radiological Sciences and Public Health, University of Brescia, Brescia, Italy.

David A. Wood, Centre for Heart Valve Innovation, St. Paul’s and Vancouver General Hospital, Vancouver, BC, Canada. Centre for Heart Valve Innovation, St. Paul’s and Vancouver General Hospital, Vancouver, BC, Canada.

Jan-Malte Sinning, Department of Cardiology, Heart Center Bonn, University Hospital Bonn, Bonn, Germany. Department of Cardiology, Heart Center Bonn, University Hospital Bonn, Bonn, Germany.

Jan Van Der Heyden, Sint-Jan Hospital, Bruges, Belgium. Sint-Jan Hospital, Bruges, Belgium.

Dirk-Jan van Ginkel, Department of Cardiology, St Antonius Hospital, Nieuwegein, the Netherlands. Department of Cardiology, St Antonius Hospital, Nieuwegein, the Netherlands.

Nicholas Van Mieghem, Thoraxcenter, Erasmus University Medical Center, Rotterdam, the Netherlands. Thoraxcenter, Erasmus University Medical Center, Rotterdam, the Netherlands.

Verena Veulemans, Division of Cardiology, Pulmonology, and Vascular Medicine, University Hospital Düsseldorf, Düsseldorf, Germany. Division of Cardiology, Pulmonology, and Vascular Medicine, University Hospital Düsseldorf, Düsseldorf, Germany.

Darren Mylotte, Department of Cardiology, Galway University Hospital, Galway, Ireland. Department of Cardiology, Galway University Hospital, Galway, Ireland.

Vasileios Tzalamouras, Cardiology, King's College Hospital, London, UK. Cardiology, King's College Hospital, London, UK.

Maurizio Taramasso, Division of Cardiothoracic Surgery, Arzt bei HerzZentrum Hirslanden Zürich, Zürich, Switzerland. Division of Cardiothoracic Surgery, Arzt bei HerzZentrum Hirslanden Zürich, Zürich, Switzerland.

Rodrigo Estévez-Loureiro, Cardiology Department, University Hospital Alvaro Cunqueiro, Galicia Sur Health Research Institute, Vigo, Spain. Cardiology Department, University Hospital Alvaro Cunqueiro, Galicia Sur Health Research Institute, Vigo, Spain.

Antonio Colombo, Cardio Center, Humanitas Research Hospital, Rozzano-Milan, Italy. Cardio Center, Humanitas Research Hospital, Rozzano-Milan, Italy.

Antonio Mangieri, Cardio Center, Humanitas Research Hospital, Rozzano-Milan, Italy. Cardio Center, Humanitas Research Hospital, Rozzano-Milan, Italy.

Azeem Latib, Montefiore Einstein Center for Heart and Vascular Care, Montefiore Medical Center, Albert Einstein College of Medicine, Bronx, NY, USA. Montefiore Einstein Center for Heart and Vascular Care, Montefiore Medical Center, Albert Einstein College of Medicine, Bronx, NY, USA.

References

  • Writing Committee, Otto CM, Nishimura RA, Bonow RO, Carabello BA, Erwin JP, Gentile F, Jneid H, Krieger EV, Mack M, McLeod C, O'Gara PT, Rigolin VH, Sundt TM, Thompson A, Toly C. 2020 ACC/AHA Guideline for the Management of Patients With Valvular Heart Disease: A Report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. J Am Coll Cardiol. 2021;77:e25–197. [PubMed] [Google Scholar]
  • Vahanian A, Beyersdorf F, Praz F, Milojevic M, Baldus S, Bauersachs J, Capodanno D, Conradi L, De Bonis, De Paulis, Delgado V, Freemantle N, Haugaa KH, Jeppsson A, Jüni P, Pierard L, Prendergast BD, Sádaba JR, Tribouilloy C, Wojakowski W. 2021 ESC/EACTS Guidelines for the management of valvular heart disease. EuroIntervention. 2022;17:e1126–96. [PMC free article] [PubMed] [Google Scholar]
  • Athappan G, Patvardhan E, Tuzcu EM, Svensson LG, Lemos PA, Fraccaro C, Tarantini G, Sinning JM, Nickenig G, Capodanno D, Tamburino C, Latib A, Colombo A, Kapadia SR. Incidence, predictors, and outcomes of aortic regurgitation after transcatheter aortic valve replacement: meta-analysis and systematic review of literature. J Am Coll Cardiol. 2013;61:1585–95. [PubMed] [Google Scholar]
  • Pagnesi M, Kim WK, Conradi L, Barbanti M, Stefanini GG, Zeus T, Pilgrim T, Schofer J, Zweiker D, Testa L, Taramasso M, Hildick-Smith D, Abizaid A, Wolf A, Van Mieghem, Sedaghat A, Wöhrle J, Khogali S, Van der, Webb JG, Estévez-Loureiro R, Mylotte D, MacCarthy P, Brugaletta S, Hamm CW, Bhadra OD, Schäfer U, Costa G, Tamburino C, Cannata F, Reimers B, Veulemans V, Asami M, Windecker S, Eitan A, Schmidt A, Bianchi G, Bedogni F, Saccocci M, Maisano F, Alsanjari O, Siqueira D, Jensen CJ, Naber CK, Ziviello F, Sinning JM, Seeger J, Rottbauer W, Brouwer J, Alenezi A, Wood DA, Tzalamouras V, Regueiro A, Colombo A, Latib A. Transcatheter Aortic Valve Replacement With Next-Generation Self-Expanding Devices: A Multicenter, Retrospective, Propensity-Matched Comparison of Evolut PRO Versus Acurate neo Transcatheter Heart Valves. JACC Cardiovasc Interv. 2019;12:433–43. [PubMed] [Google Scholar]
  • Kim WK, Möllmann H, Liebetrau C, Renker M, Rolf A, Simon P, Van Linden, Arsalan M, Doss M, Hamm CW, Walther T. The ACURATE neo Transcatheter Heart Valve: A Comprehensive Analysis of Predictors of Procedural Outcome. JACC Cardiovasc Interv. 2018;11:1721–9. [PubMed] [Google Scholar]
  • Pagnesi M, Kim WK, Conradi L, Barbanti M, Stefanini GG, Schofer J, Hildick-Smith D, Pilgrim T, Abizaid A, Zweiker D, Testa L, Taramasso M, Wolf A, Webb JG, Sedaghat A, Van der, Ziviello F, MacCarthy P, Hamm CW, Bhadra OD, Schäfer U, Costa G, Tamburino C, Cannata F, Reimers B, Eitan A, Alsanjari O, Asami M, Windecker S, Siqueira D, Schmidt A, Bianchi G, Bedogni F, Saccocci M, Maisano F, Jensen CJ, Naber CK, Alenezi A, Wood DA, Sinning JM, Brouwer J, Tzalamouras V, Van Mieghem, Colombo A, Latib A. Impact of Predilatation Prior to Transcatheter Aortic Valve Implantation With the Self-Expanding Acurate neo Device (from the Multicenter NEOPRO Registry). Am J Cardiol. 2020;125:1369–77. [PubMed] [Google Scholar]
  • Mangieri A, Chieffo A, Kim WK, Stefanini GG, Rescigno G, Barbanti M, Tamburino C, Rück A, Pagnesi M, Linder R, Toggweiler S, Montorfano M, Colombo A. Transcatheter aortic valve implantation using the ACURATE neo in bicuspid and tricuspid aortic valve stenosis: a propensity-matched analysis of a European experience. EuroIntervention. 2018;14:e1269–75. [PubMed] [Google Scholar]
  • Kim WK, Hengstenberg C, Hilker M, Kerber S, Schäfer U, Rudolph T, Linke A, Franz N, Kuntze T, Nef H, Kappert U, Zembala MO, Toggweiler S, Walther T, Möllmann H. The SAVI-TF Registry: 1-Year Outcomes of the European Post-Market Registry Using the ACURATE neo Transcatheter Heart Valve Under Real-World Conditions in 1,000 Patients. JACC Cardiovasc Interv. 2018;11:1368–74. [PubMed] [Google Scholar]
  • Kim WK, Hengstenberg C, Hilker M, Schäfer U, Rudolph TK, Toggweiler S, Rück A, Søndergaard L, Conradi L, Hamm C, Walther T, Möllmann H Collaborators. Transcatheter aortic valve implantation with the ACURATE neo valve: Indications, procedural aspects and clinical outcomes. EuroIntervention. 2020;15:e1571–9. [PubMed] [Google Scholar]
  • Tamburino C, Bleiziffer S, Thiele H, Scholtz S, Hildick-Smith D, Cunnington M, Wolf A, Barbanti M, Tchetchè D, Garot P, Pagnotta P, Gilard M, Bedogni F, Van Belle, Vasa-Nicotera M, Chieffo A, Deutsch O, Kempfert J, Sondergaard L, Butter C, Trillo-Nouche R, Lotfi S, Möllmann H, Joner M, Abdel-Wahab M, Bogaerts K, Hengstenberg C, Capodanno D. Comparison of Self-Expanding Bioprostheses for Transcatheter Aortic Valve Replacement in Patients With Symptomatic Severe Aortic Stenosis: SCOPE 2 Randomized Clinical Trial. Circulation. 2020;142:2431–42. [PubMed] [Google Scholar]
  • Lanz J, Kim WK, Walther T, Burgdorf C, Möllmann H, Linke A, Redwood S, Thilo C, Hilker M, Joner M, Thiele H, Conzelmann L, Conradi L, Kerber S, Schymik G, Prendergast B, Husser O, Stortecky S, Heg D, Jüni P, Windecker S, Pilgrim T SCOPE I investigators. Safety and efficacy of a self-expanding versus a balloon-expandable bioprosthesis for transcatheter aortic valve replacement in patients with symptomatic severe aortic stenosis: a randomised non-inferiority trial. Lancet. 2019;394:1619–28. [PubMed] [Google Scholar]
  • Rück A, Kim WK, Kawashima H, Abdelshafy M, Elkoumy A, Elzomor H, Wang R, Meduri CU, Verouhis D, Saleh N, Onuma Y, Mylotte D, Serruys PW, Soliman O. Paravalvular aortic regurgitation severity assessed by quantitative aortography: ACURATE neo2 versus ACURATE neo Transcatheter Aortic Valve Implantation. J Clin Med. 2021;10:4627. [PMC free article] [PubMed] [Google Scholar]
  • VARC-3 WRITING, Généreux P, Piazza N, Alu MC, Nazif T, Hahn RT, Pibarot P, Bax JJ, Leipsic JA, Blanke P, Blackstone EH, Finn MT, Kapadia S, Linke A, Mack MJ, Makkar R, Mehran R, Popma JJ, Reardon M, Rodes-Cabau J, Van Mieghem, Webb JG, Cohen DJ, Leon MB. Valve Academic Research Consortium 3: Updated Endpoint Definitions for Aortic Valve Clinical Research. J Am Coll Cardiol. 2021;77:2717–46. [PubMed] [Google Scholar]
  • Barbanti M, Yang TH, Rodès Cabau, Tamburino C, Wood DA, Jilaihawi H, Blanke P, Makkar RR, Latib A, Colombo A, Tarantini G, Raju R, Binder RK, Nguyen G, Freeman M, Ribeiro HB, Kapadia S, Min J, Feuchtner G, Gurtvich R, Alqoofi F, Pelletier M, Ussia GP, Napodano M, De Brito, Kodali S, Norgaard BL, Hansson NC, Pache G, Canovas SJ, Zhang H, Leon MB, Webb JG, Leipsic J. Anatomical and procedural features associated with aortic root rupture during balloon-expandable transcatheter aortic valve replacement. Circulation. 2013;128:244–53. [PubMed] [Google Scholar]
  • Möllmann H, Holzhey DM, Hilker M, Toggweiler S, Schäfer U, Treede H, Joner M, Søndergaard L, Christen T, Allocco DJ, Kim WK. The ACURATE neo2 valve system for transcatheter aortic valve implantation: 30-day and 1-year outcomes. Clin Res Cardiol. 2021;110:1912–20. [PMC free article] [PubMed] [Google Scholar]
  • Regazzoli D, Chiarito M, Cannata F, Pagnesi M, Miura M, Ziviello F, Picci A, Reifart J, De Marco, Bedogni F, Adamo M, Curello S, Teles R, Taramasso M, Barbanti M, Tamburino C, Stefanini GG, Mangieri A, Giannini F, Pagnotta PA, Maisano F, Kim WK, Van Mieghem, Colombo A, Reimers B, Latib A TAVI-SMALL Investigators. Transcatheter Self-Expandable Valve Implantation for Aortic Stenosis in Small Aortic Annuli: The TAVI-SMALL Registry. JACC Cardiovasc Interv. 2020;13:196–206. [PubMed] [Google Scholar]
  • Abdelghani M, Mankerious N, Allali A, Landt M, Kaur J, Sulimov DS, Merten C, Sachse S, Mehilli J, Neumann FJ, Frerker C, Kurz T, El-Mawardy M, Richardt G, Abdel-Wahab M. Bioprosthetic Valve Performance After Transcatheter Aortic Valve Replacement With Self-Expanding Versus Balloon-Expandable Valves in Large Versus Small Aortic Valve Annuli: Insights From the CHOICE Trial and the CHOICE-Extend Registry. JACC Cardiovasc Interv. 2018;11:2507–18. [PubMed] [Google Scholar]
  • Mauri V, Kim WK, Abumayyaleh M, Walther T, Moellmann H, Schaefer U, Conradi L, Hengstenberg C, Hilker M, Wahlers T, Baldus S, Rudolph V, Madershahian N, Rudolph TK. Short-Term Outcome and Hemodynamic Performance of Next-Generation Self-Expanding Versus Balloon-Expandable Transcatheter Aortic Valves in Patients with Small Aortic Annulus: A Multicenter Propensity-Matched Comparison. Circ Cardiovasc Interv. 2017;10:e005013. [PubMed] [Google Scholar]
  • Popma JJ, Deeb GM, Yakubov SJ, Mumtaz M, Gada H, O’Hair D, Bajwa T, Heiser JC, Merhi W, Kleiman NS, Askew J, Sorajja P, Rovin J, Chetcuti SJ, Adams DH, Teirstein PS, Zorn GL, Forrest JK, Tchetche D, Resar J, Walton A, Piazza N, Ramlawi B, Robinson N, Petrossian G, Gleason TG, Oh JK, Boulware MJ, Qiao H, Mugglin AS, Reardon MJ Evolut Low Risk Trial Investigators. Transcatheter Aortic-Valve Replacement with a Self-Expanding Valve in Low-Risk Patients. N Engl J Med. 2019;380:1706–15. [PubMed] [Google Scholar]
  • Nai Fovino, Cipriani A, Fabris T, Massussi M, Scotti A, Lorenzoni G, Guerra MC, Cardaioli F, Rodinò G, Matsuda Y, Masiero G, Leoni L, Zorzi A, Fraccaro C, Tarantini G. Anatomical Predictors of Pacemaker Dependency After Transcatheter Aortic Valve Replacement. Circ Arrhythm Electrophysiol. 2021;14:e009028. [PubMed] [Google Scholar]

Articles from EuroIntervention are provided here courtesy of Europa Group

Haemodynamic performance and clinical outcomes of transcatheter aortic valve replacement with the self-expanding ACURATE neo2 (2024)

FAQs

Which is better balloon expandable or self expanding transcatheter aortic valve replacement? ›

Among patients with severe aortic stenosis and a small aortic annulus who underwent TAVR, a self-expanding supraannular valve was noninferior to a balloon-expandable valve with respect to clinical outcomes and was superior with respect to bioprosthetic-valve dysfunction through 12 months.

What is the success rate of a transcatheter aortic valve replacement? ›

Long-term outcomes

The participants in the study were followed up for a median duration of 5.2 years, and the interquartile range (IQR) extended from 4.3 to 6.8 years. Overall, the long-term survival rate after TAVR was 73.5% (95% CI: 69.8%-77.1%).

What to expect after transcatheter aortic valve replacement? ›

Transcatheter Aortic Valve Replacement Recovery

You'll likely spend one or two days in the hospital following a TAVR procedure. You may need to continue taking certain medications, such as blood thinners to prevent clots, after aortic valve replacement. You'll need to avoid strenuous activities for one to two weeks.

What are the outcomes of TAVI? ›

TAVI device failure management

Surgical explantation was associated with a remarkable risk of early death (15–30% at 30 days), probably related to the very sick patient population, the limited surgical expertise and the need for concomitant procedures, such as root replacement and enlargement9.

What is one of the main advantages of the transcatheter aortic valve replacement TAVR procedure? ›

Some of those advantages include: Less invasive than surgery. Most surgeries on your heart involve larger incisions in your chest. Surgeons also often have to lift — or even crack and spread open — your rib cage to do heart surgery.

Which heart valve is hardest to replace? ›

The tricuspid valve is even more challenging: a leaking tricuspid valve is hard to treat both with surgery and with percutaneous treatments. In terms of recovery, open heart surgery will generally mean a one-week stay in hospital and four-to-six weeks of recovery at home.

What is the downside of TAVR? ›

Risks of TAVR

Still, the procedure carries some risk, which our team thoroughly explains beforehand and works hard to avoid: Valve leaks: Sometimes blood leaks around the new valve because the replacement is not big enough, did not fully expand, or has interference from calcium buildup.

What is the life expectancy of a person on TAVR? ›

After TAVR, the 5-year survival of patients with low (78 ± 8 years) or intermediate (83 ± 5 years) surgical risk was significantly lower than that of the general population (low risk, 64 % vs.

What are the lifestyle changes after TAVR? ›

Limitations after TAVR for most people include no heavy lifting over five to 10 lbs. for one week. Patients may not drive for a week and should not submerge the incision site(s) in water, such as a bathtub, swimming pool or hot tub. Showering is allowed.

What are the big five complications after a transcatheter aortic valve replacement? ›

Nonetheless, we have learned over the years that 5 periprocedural complications may occur that impact survival after TAVR more or less significantly: moderate/severe paravalvular leakage (PVL) (5,6); major vascular and bleeding complications (7,8); disabling stroke (9); acute kidney injury (AKI) (10); and conduction ...

Can you live 20 years after aortic valve replacement? ›

The study showed that people with aortic stenosis who had surgery to replace a narrowed valve have only a slightly reduced life expectancy than people without the condition. Post-operative survival ranged from an average of 16 years in people aged 65 or less, to between six and seven years for people over 75.

What is the most common complication of TAVI? ›

TAVI can damage the heart's electrical system, slowing down the heart and resulting in the need for a permanent pacemaker. This is the most common complication of TAVI, affecting between 1 in 20 and 1 in 10 people. heart rhythm abnormalities: there is a risk of arrhythmias following the procedure.

What is the quality of life after a TAVI? ›

Conclusions. TAVI is associated with significant early improvement in QOL, which is sustained at 1 year. The inclusion of QOL can support treatment decisions and patient-centred evaluation.

What is the mortality rate for TAVI patients? ›

The overall survival in the entire TAVI population was 58% at 5 years and 20% at 10 years (Figure 1C).

Which is better balloon expandable or self expandable stents? ›

Accordingly, BX stents resist the balloon expansion process, whereas SX stents assist vessel expansion. BX stents were the first to be commercially distributed and, even though SX stents have captured a significant share of the peripheral vascular market, BX stents remain dominant in the much larger coronary arena.

What is the most common approach for transcatheter aortic valve replacement TAVR )? ›

A TAVR using the femoral artery, a blood vessel in your groin or thigh, is the most common. Your doctor guides a tube with the replacement valve through the femoral artery to the heart. Blood vessels in the chest may be used by the doctor to guide the tube to your heart if your femoral artery is too small or damaged.

What is the best aortic valve replacement? ›

TAVR, or Transcatheter Aortic Valve Replacement, is quickly becoming the new gold-standard for treating people suffering from aortic valve disease.

What are the advantages of self-expanding stents? ›

As the sheath or wire is removed, the stent expands on its own and conforms to the shape of the artery. One advantage of self-expanding stents is that they can treat blood vessels in difficult-to-reach locations, such as the brain or the aorta, where balloon-expandable stents may not be suitable.

Top Articles
'A nice welcome back': Spencer Torkelson returns with two big hits as Tigers top Yankees
Stau in Hessen: Aktuelle Verkehrsmeldungen
LAC-318900 - Wildfire and Smoke Map
Atrium Attorney Portal
A Comprehensive Guide to Redgif Downloader
U-Bolts - Screws, Bolts variety of type & configurable | MISUMI Thailand
Update | Een maand afvallen met NBFM (+ recept & snacktips!) - Mama's Meisje
Sinai Web Scheduler
Blaire White's Transformation: Before And After Transition
Rogers Breece Obituaries
Elisabeth Fuchs, Conductor : Magazine : salzburg.info
Cappacuolo Pronunciation
Sitcoms Online Message Board
Zipcar Miami Airport
Does Publix Have Sephora Gift Cards
Dtm Urban Dictionary
Ratchet And Clank Tools Of Destruction Rpcs3 Freeze
Promiseb Discontinued
Dr. Katrina (Katrina Hutchins) on LinkedIn: #dreambig #classof2025 #bestclassever #leadershipaugusta
Kay Hansen blowj*b
What is a Nutmeg in Soccer? (Explained!) - Soccer Knowledge Hub
craigslist: northern MI jobs, apartments, for sale, services, community, and events
Kristian Andersen | Scripps Research
Antonios Worcester Menu
Walgreens Pharmacy On Jennings Station Road
Joshua Efird Net Worth
Trade Chart Dave Richard
Baldurs Gate 3 Igg
How to Start a Travel Agency: Steps and Tips | myPOS
Adventhealth Employee Handbook 2022
Red Dragon Fort Mohave Az
Age Of Attila's Rain Crossword
Hd Hub4U Com
Best Truck Lease Deals $0 Down
charleston rooms & shares - craigslist
Megan Bayne Has Made A Mega Mark Since Arriving In Stardom
Rwby Crossover Fanfiction Archive
Con Edison Outage Map Staten Island
Chloe Dicarlo
600 Aviator Court Vandalia Oh 45377
Yakini Q Sj Photos
76 Games Unblocked Fnf
Nusl Symplicity Login
Norville Breast Center At Alamance Regional
Olive Onyx Amora
Poopybarbz
Six Broadway Wiki
Why Did Jen Lewis Leave Wavy 10
A Man Called Otto Showtimes Near Cinemark Palace 20
Pamibaby Telegram
Cpc 1190 Pill
Pnp Telegram Group
Latest Posts
Article information

Author: Sen. Emmett Berge

Last Updated:

Views: 5745

Rating: 5 / 5 (60 voted)

Reviews: 83% of readers found this page helpful

Author information

Name: Sen. Emmett Berge

Birthday: 1993-06-17

Address: 787 Elvis Divide, Port Brice, OH 24507-6802

Phone: +9779049645255

Job: Senior Healthcare Specialist

Hobby: Cycling, Model building, Kitesurfing, Origami, Lapidary, Dance, Basketball

Introduction: My name is Sen. Emmett Berge, I am a funny, vast, charming, courageous, enthusiastic, jolly, famous person who loves writing and wants to share my knowledge and understanding with you.